Connecting the Cylinder
Before connecting the cylinder:
1.
Make sure the UltraFill is plugged in and the concentrator is turned on.
Note: allow the concentrator to warm up for 5 minutes prior to filling a cylinder.
2.
Press the power button to turn on the UltraFill. The green, yellow and red level indicator lights will flash and the
system will beep. Then the yellow and red level indicators will turn off and the green level indicator light will be
solid.
3.
Check to see that the OCD or regulator is set to "OFF".
Note: if filling the cylinder without an OCD or regulator attached, please close the cylinder valve to ensure the cylinder fills
properly.
To connect the cylinder:
1.
Remove the protective caps from the oxygen cylinder connector and cylinder fill connector.
2.
Line up and place the cylinder fill connector inside the oxygen cylinder connector.
3.
Lightly push the cylinder in to the oxygen cylinder connector.