1
Behaviour after switch-on
For 5 s: Lamp test
For further 60 s display: "RAB"
Timeout:
No operation
in 60 minutes
"ZE"
Zero-point calibration
"SP"
Sensitivity calibration
Operation for maintenance
Display of the measured value and the configured measuring unit
short (>1 s and <3 s)
+
Initial display:
Current measured value
+
+
Initial display:
Current measured value
+
Display: "
"
or
(press any key)
Display: " PW "
+
Enter password with
or
(Acknowledge each digit entry by pressing
Password "CCC"
+
Fault has
occurred
Fault
Display "E00" to "E12"
Calibration menu
+
Calibration of
the zero-point,
page 33
+
Calibration
of sensitivity,
page 33
Standby mode
or
)
Fault
rectified
Timeout:
If no key is pressed in any
condition for longer than
4 minutes, the menu is
closed automatically without
saving and the display
switches back to operation
for maintenance.
00433190_en.eps
31