Installation on
mobile units and
baskets with silic-
one adapters
A 813 adapter
A 812 adapter
The A 814 holder must be firmly screwed in place. In the case of bas-
kets and mobile units with silicone adapters, an A 812 adapter must
be used that can then be fitted with up to 3 A 814 holders.
Inserting the A 814 holder into a silicone adapter is not permit-
ted.
Filters
When the A 814 holder is used, the wash water must be filtered.
The injector manifold must be fitted with an A 800 tubular filter.
The tubular filter is supplied with its own operating instructions.
Follow the instructions on how to perform regular cleaning on the
tubular filter.
When the A 813 adapter is used, an FP 20 filter plate can be inserted
into the adapter instead of the tubular filter.
Do not insert a filter plate into the adapter if the injector manifold
has been fitted with an A 800 tubular filter.
Injector manifolds with silicone adapters cannot be fitted with a tubu-
lar filter. The A 812 adapter must therefore always be fitted with an
FP 20 filter plate for this application. Observe the operating instruc-
tions for the adapter.
en - Areas of application
29