Cause
Has the test strip been
lying around open for
a long time?
Was the test strip
container closed
properly?
Did the monitor work
properly?
Is the control solution
soiled or has it
exceeded the expiry
date?
Were the test strips
and control solution
stored in a cool, dry
place?
Did you follow the
steps of the test
procedure properly?
5 Operation
Remedy
Repeat the test with a
test strip that has been
stored correctly.
The test strips are damp.
Replace the test strip.
Repeat the test as
described in section
5.1.2. If problems persist,
get in touch with the
service centre.
Use new control solution
to check the performance
of the measuring system.
Repeat the control test
using strips and solution
that have been stored
correctly.
Repeat the test as
described in section
5.1.2. If problems persist,
get in touch with the
service centre.
27
GB