Selecting The Radiographic Technique - Sirona HELIDENT DS Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Sirona Dental Systems GmbH
Operating Instructions HELIODENT DS
5.2

Selecting the radiographic technique

6
60 81 744 D 3302
D 3302.201.01.25.02
5.2 Selecting the radiographic technique
Digital
If you are working with a digital radiographic system (e.g.
SIDEXIS), the Digital mode indicator lamp (6) on the
unit must be lit up.
You can switch back and forth between the conventional
and the digital radiographic technique by briefly holding
down the D button. The resulting reduction in the expo-
sure time will then be displayed.
Use an X-ray field limitation for the digital radiographic
technique.
i
NOTE
To set the recommended exposure times for adults, al-
ways set the patient symbol to the highest exposure level
within the relevant tooth symbol.
If you are working with children, always set to the lowest
exposure level within the relevant tooth symbol!
Conventional
If you want to take conventional (film) exposures, the
Digital mode indicator (6) must not be lit up.
You can switch back and forth between the conventional
and digital radiographic technique by briefly pressing the
D button. The Digital mode indicator (6) then switches
off.
The exposure time for conventional exposures is then
displayed.
Insert the X-ray field limitation for the conventional radi-
ographic technique.
5 Preparing an exposure
13