This device has been validated to withstand up to 5000 cleaning
and disinfection cycles using the recommended disinfecting wipe/
towelette. The tested number of cycles is estimated by 2 cleaning and
disinfection cycles per day over 5 years, the expected life of the device.
The meter should be replaced after the validated number of cleaning
and disinfection cycles or the warranty period, which ever comes first.
Improper system cleaning and disinfection may result in meter mal-
function. Stop using the meter if you see any signs of deterioration,
for example, LCD display cracks or becomes cloudy, buttons no longer
function, or outer casing cracks. If you have any question. Please
contact local customer services for a replacement meter if any of the
signs of deterioration are noticed.
• Do NOT clean and disinfect the meter while performing tests.
• Please follow the instructions on the package label of Micro-Kill+™ disin
fecting wipes for safe use of the wipes
• If the meter is being operated by a second person, the meter and lancing
device should be decontaminated prior to use by the second person.
• Do NOT allow cleaning and disinfecting solution to get in the test slot, bat
tery compartment, or strip-ejection button.
• If you do get moisture in the test strip slot, wipe it away with a corner of
tissue.
• Always dry the meter thoroughly before using it
• Do not spray the meter directly with cleaning solutions especially those
containing water (i.e. soapy water), as this could cause the solution to enter
the case inside and damage the electronic components or circuitry
NOTE
43