4 Operation

4.1 Electrical connections

WMS weigh modules have a standard RS232C and RS422 interface for the communication with the
control systems.
Interface descriptions
Interface type
Interface update rate (max.)
Power supply for the weigh module
Input voltage
Use a stable power supply with no voltage fluctuations.
If voltage fluctuations cannot be prevented, use a voltage regulator to deliver a constant voltage
value to the load cell.
The power supply must be approved by the respective national test center of the country in which
the weigh module will be used.

4.2 Environmental conditions

WMS weigh modules can be operated within the following environmental conditions:
Temperature Range
Relative air humidity
Height above mean sea level
Warm-up time

4.3 Ingress protection

IP rating of the WMS weigh modules:
IP54: when weighing (protected with double-labyrinth).
IP66: when cleaning wash-down (seal activated with 1 bar air pressure).
Weigh Modules
RS232C, bidirectional, full duplex
RS422, bidirectional, full duplex, bus-capable
92 weight values per second
12 to 24 V DC nominal (10 - 29 V DC)
Operating /
Compensated
Allowable ambient
+10 ... +30 °C
+5 ... +40 °C
Max. 80% at 31 °C, decreasing linearly to 50% at
40 °C, non-condensing
Max. 4'000 m (13'330 ft)
At least 30 minutes after the weigh module has
been connected to the mains.