Symptom:
System will not operate: Power ON Indicator is
illuminated
1. If the office has more than one Tap-On Foot Pedal, test each
to ensure that the proper Tap-On Foot Pedal is being used.
With a handpiece and insert installed, depress the Tap-On
Foot Pedal to the first position. The system should dispense
water. If none of the Tap-On Foot Pedals operate the system,
continue to the next step.
2. Ensure the foot pedal power is ON. Resynchronize one Tap-On
Foot Pedal to the system (see Section 7.7 Tap-On Foot Pedal
Synchronization).
Symptom:
System operates: No water flow to insert tip or
handpiece overheats
1. Assure that handpiece lavage control is properly adjusted.
2. Check for clogged insert. Replace insert if necessary.
3. Check that dental office water supply valves are open.
4. If the system is connected to DualSelect Dispensing System,
check that fluid level in the selected bottle is sufficient. Make
sure valves are open when using external water source.
5. Check that the water line filter is clean. Replace filter if
needed.
Symptom:
System operates: No insert cavitation
1. Check the insert for damage and that it is properly installed in
the handpiece.
2. Check that the handpiece is properly installed to the cable
assembly.
3. Verify that the soft nozzle grip is flush with the hard plastic of
the insert port.
4. Turn the system's Main Power Switch to the OFF (0) position.
Wait 5 seconds and turn the system back ON.
5. If problem still exists, connect the Auxiliary Foot Pedal Cable.
Symptom:
System operates: Purge Mode will not function –
Service Icon 1
1. Check that handpiece is properly installed to the cable
assembly and the purge screen is visible.
Symptom:
System operates: Service Indicator Illuminated
• Service Icon 1 Illuminated – Indicates improper set-up
1. If insert is in the handpiece, remove. Verify the handpiece
is properly seated and hold the foot pedal for 2 seconds. If
icon turns off, the system is ready for use. If icon remains lit,
continue to the next step.
2. Attach a NEW handpiece and hold Foot Pedal for 2 seconds.
If icon turns off, the system is ready for use. Discard the old
handpiece or return if within warranty. If icon remains lit,
continue to the next step.
3. If icon remains lit, refer to Section 11.2 Technical Support and
Repairs to have unit serviced as soon as possible.
System operates: Service Indicator illuminated
• Service Icon 2 Illuminated
1. Ensure that the base unit has adequate ventilation and is not
near a heat source (i.e. radiator, heat lamp, sunlight or other
heat producing operatory equipment).
2. Turn Main Power Switch to the OFF (O) position. Allow system
to cool for 10 minutes and turn system to the ON (I) position.
Verify Service Icon 2 is not illuminated.
3. If icon is still illuminated, refer to section 11.2 Technical
Support and Repairs to have unit serviced as soon as possible.

11.2 Technical Support and Repairs

For technical support and repair assistance call DENTSPLY Cavitron
Care Factory Certified Service at 1-800-989-8826 Monday through
Friday, 8:00 A.M. to 5:00 P.M. (Eastern Time). For areas outside the U.S.,
contact your local DENTSPLY representative.
SECTION 12:

Warranty Period

The Cavitron Touch Ultrasonic Scaler is warranted for TWO YEARS
from date of purchase. The Steri-Mate 360 Handpiece enclosed with
your system is warranted for SIX MONTHS from date of purchase.
Refer to the Warranty Statement Sheet furnished with your system
for full Warranty Statement and Terms.