Beam Alignment Procedure
While watching the camera view on the touchscreen, use the beam alignment joystick on the
operator side of Valeda to move the targeting beam (yellow circle) le /right and up/down until the
beam fills the eye socket. The picture on the touch screen illustrates the correct position of the
beam and the patient's eye to be viewed (Figure 10).
FIGURE 10
Once the beam is aligned, instruct the patient to remain in position until instructed to move.
Press the "Confirmation"
button on the touchscreen (Figure 10).
Treatment Procedure
The treatment will consist of 4 phases:
1. The treatment initiates with 35 seconds of pulsed yellow and near-infrared (NIR) wavelengths,
eyes open.
2. This is followed by 90 seconds of continuous red light wavelength, eyes closed.
3. The treatment will repeat with another 35 seconds of pulsed yellow and NIR wavelengths, eyes
open.
4. The treatment will finish with another 90 seconds of continuous red light wavelength, eyes closed.
The operator will instruct the patient to keep their eyes open during phases 1 and 3 and then to
keep their eyes closed for phases 2 and 4.