16. Technical Specifications

Parameter
Size
Weight
Electrical Power Input
Fuse Rating
Light Sources
Light Emission
Beam Diameter
Treatment
Exposure Time
Operating
Environment
On-site Storage
Environment
Transportation
Environment
Ingress Rating
Electrical Safety
Classification
Electromagnetic
Compatibility
Photobiological
Safety
Specification
530 mm height x 300 mm width x 330 mm depth
10.8 kg
100 – 240 VAC, 50/60 Hz, 1.0 – 0.5 A
Voltage: 250 V
Current: 2 A
Operating speed: Slow
Size: 5 x 20 mm
IEC Fuse Code: T2AL250V
Light emitting diodes (LEDs)
590 nm output: 5 mW/cm2
660 nm output: 65 mW/cm2
850 nm output: 8 mW/cm2
30 mm (nominal) at treatment plane
A total of 250 seconds (4 minutes 10 seconds). There are 4 phases:
Phase 1 – 35 seconds, patient's eyes open
Phase 2 – 90 seconds, patient's eyes closed
Phase 3 – 35 seconds, patient's eyes open
Phase 4 – 90 seconds, patient's eyes closed
Temperature: 15 – 30°C
Humidity: 15 – 90% (non-condensing)
Altitude: Up to 3,000 m
Temperature: 15 – 30°C
Humidity: 15 – 90% (non-condensing)
Altitude: Up to 3,000 m
Temperature: -18 - 60°C
Humidity: 15 - 90% (non-condensing)
Pressure: 700 - 1060 hPa
EC 60529: IP2X
IEC 60601-1
Type BF applied part
IEC 60601-1-2
IEC 62471
Risk Group: Exempt
23