Guidance and manufacturer's declaration – electromagnetic immunity for EQUIPMENT and SYSTEMS that are not LIFE-
SUPPORTING
The Blood pressure monitor (BM 54) is intended for use in the electromagnetic environment specified below. The customer
or the user of Blood pressure monitor (BM 54) should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Conducted RF
3 V
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
Compliance
level
3 V
rms
rms
3 V/m
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used no closer
to any part of the Blood pressure monitor (BM 54), including cables, than the
recommended separation distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance.
80 MHz to 800 MHz
800 MHz to 2.5 GHz
129