Adjustment „empty"
During the adjustment „empty", with empty container, the probe calibrates to the ambient conditions. After
approx. 30 - 40 seconds the probe is ready.
The following values are indicated by the analogue output:
Current output
Current output inverted
Voltage output
Voltage output inverted
Adjustment with EasyTeach-Button (ETF):
By releasing the EasyTeach-Button after approx.
5 seconds, the menu item adjustment „empty" is
activated
Adjustment with EasyTeach by Wire (ETW):
By breaking the EasyTeach connection after approx.
5 seconds, the menu item adjustment „empty" is
activated
The blue ring flashes with a frequency of 2 Hz.
During the adjustment „empty"
• the binary output has no function
• the progress of the adjustment is indicated by the
level indicator LEDs
• the blue ring flashes with a higher frequency
During the
With successful
adjustment
adjustment
8 mA
8 mA
2,5 V
2,5 V
With failed
adjustment
4 mA
12 mA
20 mA
12 mA
0 V
10 V
5 V
5 V
23