If this equipment does cause harmful interference to radio or television reception,
which can be determined by turning the equipment off and on, the user is encouraged
to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which
the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Information regarding Electromagnetic Compatibility and Immunity
CervicalStim complies with IEC 60601-1-2 for electromagnetic compatibility (EMC).
CervicalStim needs special precautions regarding EMC and needs to be used
in accordance with the EMC information provided in this manual. Wireless
communications equipment such as home network devices, mobile phones,
cordless telephones and their base stations, and walkie-talkies can affect
CervicalStim. These types of equipment should be kept at least 0.198 m (7.8 in)
away from CervicalStim.
CervicalStim has not been evaluated with regard to use with specific implantable
electronic medical devices. Please consult your physician prior to use of the
CervicalStim with implantable electronic medical devices.