Power Supply

The device can be operated through mains operation or off line with the built-in Li-ion bat-
tery. For mains operation, the mains adapter must be connected to a power outlet and be
inserted into the left side of the device (J) using the cable. Then the battery is also char-
ged. The yellow charge indicator (F) on the front of the device is illuminated while the de-
vice is being charged. If it is not illuminated, the battery is fully charged.
For the measurement, the device must additionally be turned on on the front of the device
using button (E). As long as the On button is pressed, all segments of the display are acti-
vated. Thanks to this, segments that have failed can be clearly identified.
If no SpO2 measurement is taken and no blood pressure measurement is started either,
then the LCD illumination turns itself off in about 10 minutes. After another 5 minutes, the
entire devices turns itself off. This prevents accidental depletion of the battery.
In battery operation, the device shows the remaining charge with a symbol (9) in the lower
part of the display. This display has a large tolerance because various parameters, such as
environmental temperature, also affect the battery.

Basic Settings

The settings described below are continuously saved, even if the device is not used for a
longer period of time. The settings are done in a special mode. If button (16) is also pres-
sed when turning the device on, then you are taken to the named menu.
The following presettings are possible here:
The inflation pressure of the device is set by pressing button (13). Then current inflation
pressure is shown on the display. Independent of this setting, the device will automatically
inflate more if necessary.
The pulse signal can be turned on and off by pressing button (15). The current setting can
be seed on the signal display (8): the pulse signal is activated when the "Signal" symbol is
visible.
Press button (16) to switch the operating mode of the blood pressure measurement bet-
ween adult and neonate mode. The current setting is displayed at (11). The "NEO" setting
may only be used for neonates and infants wtube arm measures less than 13 cm in cir-
cumference.
The pulse signal and the mode setting of the blood pressure setting can also be set later
with the corresponding buttons while the device is being used. The permanent presetting
is only used for lighter operation in certain surroundings, e.g. the inflation pressure for pa-
tients who are hypertonic can be set higher from the outset.