Operation; Precautions For Use; Ratio Between Distance From Target/Diameter Of Measurement Field - AEMC CA879 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

4.1 Precautions for use

If the surface to be measured is ≤ Ø 1.96" (50mm), hold the probe as close as
possible to the surface (less than 15" [40cm] away) If the surface to be measured
is covered in frost or other particles, clean it so that it is properly exposed, before
taking a measurement.
If the surface to be measured is highly reflective, apply matt tape or paint to the
surface before measuring. If the thermometer appears to be giving incorrect indi-
cations, check the condition of the lens at the measurement opening, as it may
be obscured by condensation or particles. If so, clean it by carefully following the
instructions in § 5.
4.2 Ratio Between Distance from Target/Diameter
of Measurement Field
This ratio, also called the field of vision, indicates the diameter of the
measurement field at a given distance from the target;
D/S = 12:1 (D = distance, S = spot).
When measuring smaller areas, it is important to bring the probe sufficiently
close to the target to avoid including other points in the measurement field.
Refer to the diagram on the instrument's side panel as shown below.
8
CHAPTER 4

OPERATION

3.0@36
1.0@12
25@300
75@900
5.0@60 inch
125@1500 mm
Infrared Thermometer Model CA879

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido