8.
This Electronic Sphygmomanometer is designed for adults and should never be used for infants
or young children. Consult your physician or other health care professionals before use on older
children.
9.
Do not use this unit in a moving vehicle, This may result in erroneous measurement.
10. Blood pressure measurements determined by this monitor are equivalent to those obtained by a
trained observer using the cuff/stethoscope auscultation method, within the limits prescribed by the
American National Standard Institute, Electronic or automated sphygmomanometers.
11. If an irregular heartbeat (IHB) from common arrhythmias is detected in the procedure of blood pres-
sure measurement, this sign
nometers can keep function, but the results may not be accurate, it' s suggested that you consult with
your physician for accurate assessment.
There are 2 conditions under which the signal of IHB will be displayed:
1)
The coefficient of variation (CV) of the pulses is > 25 %.
2)
The deviation of the following pulse period is ≥ 0.14 s, and the number of such pulses amounts
to more than 53 % of the total number of measured pulses.
12. Please do not use the cuff other than supplied by the manufacturer, otherwise it may bring biocom-
patible hazard and might result in measurement error.
13.
The monitor might not meet its performance specifications or cause a safety hazard if stored or
used outside the temperature and humidity ranges specified in the specifications.
14.
Please do not share the cuff with other contagious persons to avoid cross-infection.
15. This equipment has been tested and found to comply with the limits for a Class B digital device, pursu-
ant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment generates, uses and can radiate radio
will be displayed. Under this condition, the Electronic Sphygmoma-