program has finished.
After testing all the temperatures programmed in the heat-
ing block, „tESt" is shown in the upper display and „End"
in the lower display. The heating block must then be swit-
ched off.
Safety Notice:
The temperature sensor is hot!
When removing the sensor from the heating block, grasp
the black insulation.

2.2.2 Calibration (CAL)

All temperatures programmed in the heating block are
measured by T-Set, from the lowest to the highest, regis-
tered and then saved in the heating block (this takes about
40 min). When the heating block has reached a program-
med setpoint temperature (shown in the upper display) and
calibrated the value, a signal tone is intitated. The actual
temperature measured and calibrated by T-Set is shown in
the lower display and saved in the heating block.
While the test program is running, the PROG, WAIT and
RUN LED's flash and symbolise the on-going temperature
response. The temperatures measured and calibrated by
T-Set can be logged by hand during calibration, or output
and documented with the T-Set software after the calibra-
tion has finished.
After calibration of all temperatures programmed in the
heating block, „CAL" is shown in the upper display and
„End" in the lower display. The heating block must then be
switched off.
Safety Notice:
The temperature sensor is hot!
When removing the sensor from the heating block, grasp
the black insulation.