PROBLEM
Leakage of powder from the cap
of the container.
Treatment ineffective
Apparatus in accordance with EEC Directive 93/42: Class II a.
Apparatus for intermittent operation:
Standard water pressure provided by the dental unit for a turbine:
Standard air pressure provided by the dental unit for a turbine:
Operating conditions:
Transport and storage conditions:
Weight and size:
34
POSSIBLE CAUSE
Cap not correctly screwed in
place.
Residues of powder in the thread.
Insufficient pressure of the
air-supply circuit.
Level of powder inside the
container too low or too high.
Unsuitable powder.
Front terminal of the polisher
clogged by too much moister in
the powder or insufficient clean-
ing/maintenance.
11.0 Technical specifications
60" ON 30" OFF.
From 0 to 3 bars (from 0 hPa to 3000 hPa), max
40°C.
From 2.3 to 3.2 bars (2300 hPa - 3200 hPa).
from +10°C to +40°C.
Relative humidity from 30% to 75%.
from -10°C to +70°C.
Relative humidity from 10% to 90%.
Air pressure P: 500hPa/1060hPa.
125 g.
easyjet pro - manual of use and maintenance
SOLUTION
Screw the cap on correctly.
Read point "06.1"
Check the pressure in the air
circuit of the dental unit.
Restore the correct level of the
powder inside the container
(see point "05.2").
To enable the apparatus to work
correctly use only powder
produced by MECTRON.
Read point "06.1" and "06.3".
Remove the powder from the
container and clean it with a dry
cloth.
Read point "07.0".