Functional description (cont'd)
Function generator
(cont'd)
BNC «TTL OUT»
connector
40
In "LOGIC" mode, the "LOGIC HIGH" and "LOGIC LOW" keys, when
selected (LED on), are used to change the high (HIGH) and low (LOW)
levels of the logic signal of +10 V to -10 V by steps of 100 mV.
• Adjustment is performed by the encoder.
G
These keys are active only when the "LOGIC" mode has been
selected.
On this BNC, a TTL-level logic signal is available.
• The period of these pulses is that of the signal generated at the output
"MAIN OUT", in phase and with the same duty cycle (values indicated
on the LCD).
• Adjustment of the frequency using the "FREQ" key.
• Adjustment of the duty cycle using the "DUTY" key.
G
The "AMP LEVEL" and "OFFSET" keys have no effect on the
TTL signal.
Functional description
Function generator