5.0  FUNCTIONS 
5.1  PERCENTAGE
The  balance  will  allow  a  reference  weight  to  be  shown  as  100%.  Then  any  other 
weight  placed  on  the  balance  will  be  displayed  as  a  percentage  of  the  original 
sample.  
1) Place your weight on the balance. 
2) Press the [%] key. The weight will be displayed as 100.00%.  
3) Remove  the  weight  and  place  your  next  sample  on  the  balance.  The  new 
reading will be shown as a percentage of the first value / reference used.  
4) Pressing the [%] key again will return the balance to normal weighing. 
Note:  The  balance  may  jump  by  large  numbers  unexpectedly  if  small  weights  are 
used  to  set  the  100%  level.  For  example,  if  only  23.5g  is  on  a  balance  with  0.5g 
increments and the balance is set to 100%, the display will show 100.00%. However, 
a small change of weight will cause the display to jump to 102.13%, as one balance 
division (0.5g) increase to 24.0g will be equivalent to a 2.13% increase. 
5.2  PARTS
COUNTING 
 
NTEP APPROVED NOTE: COUNTING FEATURE IS NOT LEGAL FOR TRADE 
Parts counting lets you count small parts that are of equal weight, quickly and easily. 
To do this you must first set your sample by telling the balance how many parts you 
have.  
1) Place your container on the balance and press the [‐>0/T<‐] key as described 
in  4.2  Zeroing  /  Tare  section.  This  will  remove  the  weight  of  the  container 
from your count.  
2) When the balance is in normal weighing mode with the initial quantity on the 
balance, press the [Smpl] key to start the parts counting function.  
3) The  initial  number  of  samples  should  match  the  options  for  parts  counting, 
10, 20, 50, 100 or 200 pieces. 
© Adam Equipment Company 2018