DC Accuracy
Use this procedure to verify the probe's DC accuracy. Before
beginning, read the procedure through completely.
1. Connect the test equipment as shown in Figure 4- - 3.
2. Attach the probe to a TEKPROBE interface on the 1103 power
supply.
3. Set the 1103 offset VAR/0V to 0V.
4. Set the multimeter to read DC volts.
5. Connect the probe tip and ground to the power supply using the
Y-lead and KlipChip adapters.
6. Allow 30 minutes for the equipment to warm up.
Digital Multimeter
+
- -
BNC to Dual
Banana Adapter
Figure 4- 3: P6245 DC Accuracy Setup
7. Set the power supply to +2.000 V. Record the power supply
output as Vmax.
8. Measure and record the multimeter reading as M1.
9. Set the power supply to - - 2.000 V. Record the power supply
output as Vmin.
P6245 Instruction Manual
50Ω Precision
Terminator
Performance Verification
1103
Power Supply
Power Supply
- -
+
4- 7