Table of Contents
Startup
on page 15
Specifications
Specifications are subject to change without notice.
Specification
Dimensions (W x D x L)
Enclosure
Weight
Pollution degree
Installation category
Power requirements
Fuse (AC model)
Operating conditions
Storage conditions
Data storage
Communication
Display (optional)
Accuracy
Analog outputs
Relay
Output power
Analog input for data logging 0 to +5 VDC from an external device
Setup/Data retrieval
Warranty
Details
257.8 x 95.3 x 236.7 mm (10.15 x 3.75 x 9.2 in.)
ABS plastic
2.3 kg (5 lb)
2
II
AC: 100-230 VAC, 50/60 Hz, 22/14 mA (with all loads used)
DC: 12 VDC without display, 12 VDC with display and the backlight off, 180 mA
(2.1 watts) with one 4-20 mA output used
One 5 x 20 mm, 1 A, 250 VAC, antisurge
Temperature: -20 to 50 °C (-4 to 122 °F) without display, -10 to 50 °C (14 to
122 °F) with display
Maximum relative humidity 80% for temperatures up to 31 °C (88 °F) decreasing
linearly to 50% relative humidity at 50 °C (122 °F).
Altitude: 2000 m (6,560 ft) maximum
-40 to 60° C (-40 to 140 °F) without display, -20 to 60° C (4 to 140 °F) with
display
64 kB (16 k cycles of velocity/level data)
RS232C at 19.2 K Baud
25.4 x 76.2 mm (1 x 3 in.), four lines of text
The data shown on the display is configurable. By default, the display shows the
flow, velocity, level, totalized flow and other channels.
1 second per day
Four 4-20 mA outputs, system isolated, 600 ohm load maximum
22-18 AWG wire
One dry relay contact, single pole, normally open pole contact
1 A @ 30 VDC (resistive) maximum
22-18 AWG wire
12 VDC, 150 mA to external devices
Flo-Ware for Windows software
EU: 2 years, ROW: 1 year
English 3