Instructions To Navigate Micro-Smart Menus And To Modify Some Parameters - Cattani MICRO SMART Manual De Istrucciones

Ocultar thumbs Ver también para MICRO SMART:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

• Instructions to navigate Micro-Smart
menus and to modify some parameters
Main menus
When Micro-Smart is switched on, the display shows the
Cattani logo for 10 seconds, after which time the main
menu appears.
Main Menu "A1"
This menu shows some parameters such as: suction mo-
tor speed, suction activation time, temperature, presence/
absence of the amalgam container and system software
release.
Pressing the key
takes you through
Control Menu "A2"
The display shows the number of times Micro-Smart has
been switched on by means of the main switch (Power
Cycles), the total hours Micro-Smart has had power sup-
plied (Uptime), the number of times suction has been acti-
vated (Work Cycles), the total running hours of the aspirator
(Work Time), the working hours as regards to the cycles
(HZ) used and the number of times the control panel coo-
ling fan has been activated (Fan Cycles).
Right arrow
Events Menu "A3"
This menu shows the last events or alarms that have occur-
red to Micro-Smart. Events are indicated with a number;
refer to the table at page 47 to identify them.
Control Menu "A2"
COUNTERS – ODOM . A2
POWER CYCLES
UPTIME [ h]
WORK CYCLES
WORK TIME [ h]
ASPIRATOR [ h]
FAN CYCLES
000000
Number of times the aspirator has been switched on
by means of the main switch.
000000
Total hours Micro-Smart has had power supplied
(motors switched-off).
000000
Number of times the aspirator has been activated by
the dental unit.
000000
Number of real running hours (motors running).
000000
Working hours as regards to the cycles (HZ) used.
000000
Number of times the control panel cooling fan has
been activated.
33