Table 3 - Proximity Fields from RF Wireless Communications Equipment
Test frequency
(MHz)
385
380 - 390
450
430 - 470
710
745
704 - 787
780
810
870
800 - 960
930
1720
1845
1700 - 1990
1970
2450
2400 - 2570
5240
5500
5100 - 5800
5785
PC-BMG3019_IM
Band
(MHz)
Professional healthcare facility environment
Pulse modulation 18 Hz, 27 V/m
FM, ± 5 kHz deviation, 1 kHz sine, 28 V/m
Pulse modulation 217 Hz, 9 V/m
Pulse modulation 217 Hz, 28 V/m
Pulse modulation 217 Hz, 28 V/m
Pulse modulation 217 Hz, 28 V/m
Pulse modulation 217 Hz, 9 V/m
Immunity test levels