Installation And Commissioning; Operation: Calibration Of The Electrode And Ph Transmitter - Mettler Toledo InPro 485Xi Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 18
en
16
Important! Please take note of the following technical
specifications of this pH electrode:
Reference system
Min. Na
4

Installation and commissioning

1. While unpacking, check the electrode for mechan-
ical damage. Please report any damage immedi-
ately to your METTLER TOLEDO supplier.
2. Remove the protective cap and rinse the electrode
for a short time with Buffer pH = 4.01 / Na
After rinsing, the electrode should only be lightly
dabbed. Any rubbing of the pH or the sodium-sen-
sitive glass can increase the response time as a
result of electrostatic charge.
3. Check the space behind the pH and the sodium
sensitive glass for air bubbles and remove any
air bubbles by shaking the electrode slightly in a
vertical direction.
4. Place the electrode into the housing as described
in the instructions for the housing.
5. Connect the electrode and the pH transmitter with
the corresponding cable (AK9). Observe the con-
nections scheme that accompanies the cable or
the transmitter.
6. ISM
and Measure and expanded diagnostics. For in-
stallation, commissioning and operation of the
ISM system please refer to instructions for the
transmitter, module and cable.
5
Operation: Calibration of the electrode and
pH transmitter
Before a calibration, immerse the electrode for
10 minutes in a Buffer pH = 7.00 / Na
Buffer pH = 4.01 / Na
connected to the transmitter. A 2-point calibration is
recommended, e.g. Buffer pH = 7.00 / Na
Buffer pH = 4.01 / Na
For further details please consult the operating
instructions for the pH transmitter.
InPro 485X i pH electrode
52 004 112
concentration
+
sensors such as InPro 485X i enable Plug
®
InPro 485X i pH electrode
S odium sensitive glass
membrane
10 mg / L when pH > 7;
100 mg / L when 7 > pH > 2;
1g / L
3.9M buffer solution while still
+
3.9M buffer solutions.
+
© 02 / 16 Mettler-Toledo GmbH
when pH < 2
+
3.9M or
+
3.9M and
+
Printed in Switzerland
3.9M.

Publicidad

Tabla de contenido
loading

Tabla de contenido