Ordering spare parts
When ordering spare parts, please give:
- Device type
- Manufacturing number, see type plate
- Software version
- Item number and designation of the spare part,
see www.ika.com.
Error Codes (RT 5/10/15)
Error code
Cause
Er03
Temperature inside device is too high
Er21
Safety relay doesn't open during test
Er22
Safety temperature is too low during
test
Er25
Heating and switching element
monitoring
Er26
Difference between temperature of
safety sensor and temperature of
control sensor: control temperature
> (safety temperature + 40 K)
Er31
Triac error detection doesn't work
Er46
Difference between temperature of
safety sensor and temperature of
control sensor: safety temperature
> (control temperature + 40 K)
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Please contact the service department;
- Send the device for repair, including a short description of the fault.
Accessories
• Stirring bar:
ø 8 mm; length 30 mm
• RSE
Stirring bar remover
Repair
Please send instrument in for repair only after it has been
cleaned and is free from any materials which may consti-
tute a health hazard.
For this you should request the "Safety Declaration (Deconta-
mination Certificate)" from IKA, or use the download printout
of it from the IKA website www.ika.com.
Return the instrument in its original packaging. Storage packaging
is not sufficient. Also, please use suitable shipping package mate-
rials.
Effect
Solution
Heating off
- Switch off device and allow to cool down
Heating off
- Switch off device
- Warning! Only to be carried out by authorised service personnel:
Check PCB
Heating off
- Switch off device
- Warning! Only to be carried out by authorised service personnel:
Check PCB and safety temperature sensor
Heating off
- Check the external temperature controller
- Switch off device
- Warning! Only to be carried out by authorised service personnel: Carry
out an internal test on the device to check the plug-in connector for
the heating element, the safety temperature sensor or the PCB
Heating off
- Switch off device
- Warning! Only to be carried out by authorised service personnel: Carry
out an internal test on the device to check the plug-in connector for
the temperature sensor
Heating off
- Switch off device
- Warning! Only to be carried out by authorised service personnel:
Check PCB
Heating off
- Switch off device
- Warning! Only to be carried out by authorised service personnel:
Carry out an internal test on the device to check the plug-in connec-
tor for the temperature sensor