Technical data
RO
Power supply unit
Input
Output
Protection class
Operating voltage
Power consumption (max.)
Power consumption, standby operation
Power output
Speed ranges
P50, P70, P100, ----
ECO-Mode
Set-up surface
Dimension:
Weight:
RT
Unit
Operating voltage
Frequency
Speed ranges
P50, P70, P100, ----
ECO-Mode
Power consumption (max.)
Set-up surface
Dimension:
Weight:
Adjustment and display resolution
Surface temperature
Safety temperature limit (adjustable)
Fuse:
100 V / 115 V A
RO/RT
Setting resolution
Deviation between the stirrer points
Permissible duration of operation
Permissible ambient temperature
Permissible relative humidity
EN 60529 protection class
Operation at a terrestrial altitude
Stirred quantity max. (water)
in other vessels
Stirrer points
Distance from stirrer point to stirrer point
*Caution! The specified maximum hotplate temperature refers to instruments with a full load.
Without a load, the (setup) plate temperature can reach 130 °C (+10 °C / -5 °C).
Load to determine the specified values: 250 ml Erlenmeyer flask filled with 200 ml water.
V
100 - 240
A
1,67
Vdc 24; 40 W.LPS; (limited power source)
Hz
50/60
2 (double insulated)
Vdc 24
RO 5
A
0,5
W
12
W
2
17
W
rpm 0 - 1200 in 10 rpm-steps adjustable
rpm 0 - 600 in 10 rpm-steps adjustable
RO 5
mm 120 x 450
mm 120 x 570 x 60
kg
3,0
Vac 230 ± 10%
Vac 115 ± 10%
50 / 60
Hz
rpm 0 - 1000 in 10 rpm-steps adjustable
rpm 0 - 600 in 10 rpm-steps adjustable
RT 5
W
185
mm 110 x 495
mm 120 x 610 x 60
kg
4,0
1
K
120*
°C
°C
50 - 150
RO/RT 5
230 V
A
T 2 (5x20)
T 4 (5x20)
rpm ±5
0
%
100
%
°C
+5 to +40
%
80
IP 40
m
max. 2000
400 ml per stirrer point in the 600 ml beaker glass
RO/RT 5
ltr
2
5
mm 90
RO 10
RO 15
1,0
1,5
24
36
RO 10
RO 15
180 x 450
270 x 450
190 x 570 x 60
280 x 570 x 60
4,0
7,0
RT 10
RT 15
395
600
180 x 495
270 x 495
190 x 610 x 60
280 x 610 x 60
6,5
9,4
RO/RT 10
RO/RT 15
T 4
(5x20)
T 6,3 (5x20)
T 6,3 (5x20)
T 10 (5x20)
RO/RT 10
RO/RT 15
4
6
10
15
90 x 90
90 x 90
14