Important; Scan Operating Mode - Promax 10 premium Guía De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

j)
BER RESOLUTION
LOW: BER measurements every 2s. The minimum measured BER for a
signal at 64QAM and a SR at 6875 Ks/s is 10
HIGH: BER measurements every 20 s. The minimum measured BER for a
signal at 64QAM and a SR at 6875 Ks/s is 10
CONTINUOUS: Continuous measurements up to 30 minutes. The minimum
measured BER for a signal at 64QAM and a SR at 6875 Ks/s is 10
The resolution selected in the configuration, applies also in the measurements
done in LOGGER mode, with the exception of the CONT mode that measures time of
20 s (in HIGH mode).
k) SW
The lower line of the screen shows the version of the unit control software
(7.06 in previous figure).
To leave the global configuration menu, just press the key of the operating mode
you wish to access.

4.2.2 SCAN operating mode

The SCAN operating mode shows, in a single screen, the signal level of all the
active channels in the channel plan by means of a bar-graph. In addition, the exact level
of any particular channel may be measured by simply placing the marker over it (the
measurement is calibrated only for analogue channels; for digital channels refer to the
note at the end of the section).
To
access
this
operation you should press the SCAN
[11] key. At first, the field channel (C44
in the figure) will appear shaded, indicating that is
selected and modifiable. Next to the channel, it
appears the frequency (655.25 MHz in the figure)
and the level (65 dBµV in the figure) of the channel
where the marker is pointing. The marker is moved
by using the rotary knob [9].
To
change
the
(bandwidth
represented
screen), first select it by pressing the
rotary knob until you see the "sp" shade (at the
middle bottom of the screen). Then turn the rotary
knob [9]. If you rotate clockwise the span will
increase and if you rotate anti-clockwise it will
decrease. Span can acquire the following values:
10, 30, 100, 300 MHz and full span.
Page 18
USER'S MANUAL. PROMAX-10
-7
-9
mode
of
Figure 8.- SCAN mode, full span.
span
on
the
Figure 9.- SCAN mode, span 100 MHz.
Premium
-11
.

IMPORTANT

11/2008