E
1.1.4 Septa
injection needle
Vessel is closed by a septum; this septum allows the injection of solvent and other
substances with an injection needle during work. Septum itself is made of silicone,
the inside featuring PTFE foil to ease cleaning.
1.1.5 Capacity of reaction vessels
For capacity of reaction vessels available for liquid-phase synthesis refer to list
below:
item
reaction vessel Liquid 12 552-25030-00
reaction vessel Liquid 16 552-22530-00
reaction vessel Liquid 24 552-21030-00
1.2
Locate reaction vessels in heating- and condensation block
1.2.1 Remove and install reaction vessels
Handle hot reaction block with care; danger of burns. Use protective
gloves.
P/N
total capacity
50 ml
25 ml
10 ml
64
positive lock;
holds septum in
place even under
high pressure
PTFE-foil
recommended
capacity
25 ml
12 ml
5 ml