Service Notes - Integra MAYFIELD 2 Manual De Instrucciones

Ocultar thumbs Ver también para MAYFIELD 2:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 89
Before fully tightening, always be certain that the starburst teeth of the Swivel
Adaptor and the Transitional Member fittings are the same size and properly mesh.
Failure to do so may damage fittings. Figure 8 shows a typical starburst connection
and proper meshing of teeth.

Service Notes

A routine equipment inspection made well before each case will help avoid problems. This check
should include the following:
1. Store the Skull Clamp with the Index Locking Knob in the UNLOCKED position. This practice
will provide the best performance from the knob.
2. While rotating the rocker arm through 360 degrees, lock and unlock the Index Locking Knob.
To ensure proper locking engagement, the Index Locking knob must always be turned to the
lock position. If the Index Locking Knob stops short of the locked position, a slight additional
rotation of the rocker arm should complete the internal alignment and allow the locking
mechanism to engage. If difficulties are encountered, the unit should be returned for service.
3. Check the Ratchet Extension release buttons to be sure that they allow for the Extension to
disengage freely. Press them and move the Extension out of the body portion of the clamp - do
this several times to make sure that it works smoothly and easily.
CAUTION!
Figure 8 Mounting Skull Clamp
Page 11
Proper Meshing
of Teeth

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

A3059

Tabla de contenido