Operating Instructions BA ULTIMATE POWER + / BA ULTIMATE
64 01 553 D3551.201.03.03.02
7.2

Automated cleaning and disinfecting

BA ULTIMATE POWER + / BA ULTIMATE can also be cleaned and
Product name
disinfected in suitable cleaning and disinfection equipment.
The cleaning and disinfection equipment used must be approved by its
Cleaning and disinfection equipment
manufacturer for the cleaning and disinfection of dental instruments and
comply with EN ISO 15883-1 (e.g. 95°C (203°F) and 10 min. holding
time).
For further details, refer to the operating instructions supplied with the
unit.
✔ BA ULTIMATE POWER + / BA ULTIMATE is conditioned with a
Check _ product name
cleaning and disinfection device.
1. Check whether BA ULTIMATE POWER + / BA ULTIMATE is clean
after conditioning under good lighting (min. 500 lux) and color
rendering index (min. 80 Ra).
2. If they are still dirty, repeat the process.
7 Conditioning