Display adjustment
The gage should be verified before each series of measurements by
using a gage block.
If there is a deviation from the specified measurement the device
should be re‐adjusted by using the rotating scale (E).
The measurements have to be repeated until the gage is
correctly adjusted.
In order to increase the accuracy of the measurement, the gage should be
adjusted in the same position in which later the measurement will be
effected.
Tolerance marks
The gage provides tolerance marks (D) that can be set to an acceptable tolerance range
which allows a quick measurement reading.