Standard Delivery

1
Dynamix speed mixing unit
1
Power cord
1
Instructions for use
3.4.

Accessories

Wall mounting kit
4.

Commissioning

4.1

Setup

The Dynamix speed can be used as a bench top unit, although it is also suitable for wall mounting.
4.1.1

Bench Top Unit

Select a stable underlay on which to set the unit up.
4.1.2

Wall Mounting

With the aid of the wall mounting kit (refer to the accessories), the unit can be securely mounted
on a wall. The mounting instructions are included in the mounting kit.
4.1.3

Setup Recommendations

Operate the unit at room temperature 21 – 25°C [70 – 77°F].
Depending on the information from the impression material manufacturer, the impression material is to be prepared prior
to usage in a room temperature between 21 – 25°C [70 – 77°F] for reaching optimum product properties.
Because of this, it is recommended to use Dynamix speed at the same temperature. Impression material that is too
cold, especially with temperatures less than 21°C [70°F], can result in Dynamix speed changing automatically to a lower
supporting speed because of safety reasons. Please also see chapter 9: Troubleshooting.
When setting up the unit, please note the following:
Do not place the unit under a heat source.
Do not place the unit in front of open windows.
Do not set the unit up where it will be subjected to direct sunlight.
4.2

Connection and Starting

Before connecting the unit to the wall outlet, make sure the voltage information
on the nameplate corresponds to your local power supply.
• Connect the provided power cord (19, fig. 1) to the main connection
socket (18).
• Plug the power cord into the wall outlet.
• Switch the unit on at the main power switch (16).
4.3

Remove Shipping Restraint

The shipping restraint must be removed before the unit may be operated:
• The unit has been connected and switched on as described in Section 4.2.
• Close the unit cover.
• Press the retract key (4).
○ The plungers (9, fig. 1) retract to their home position where they stop.
• Open the unit cover (8, fig. 1).
• Remove the shipping restraint (23).
The unit is now ready for use.