2. Wait only 30 seconds before taking another reading on the same patient
again to avoid excessive cooling of the skin.
3. For users in sound health, both ears are almost the same in terms of body
temperature. You should select the ear that shows a stable and relatively
higher temperature.
4. Please clean the ear before taking a reading.
NOTE: The unit may not function correctly in an environment with
electrostatic discharge so it may be necessary to reset the unit.
5.5. TROUBLES AND TROUBLESHOOTING
PROBLEMS
No response/
Automatic reset when
the insulator is taken
out
Battery symbol
on LCD
"Lo" on screen
(result below
32.0°C or 89.6°F)
"Hi" on screen
(result below
42.9°C or 109.3°F)
Errp" on screen
"ErrE" on screen
"Err" on screen

CHECK LIST

Battery used up
Battery in wrong polarity
Poor battery contact
Low battery life
Thermometer placed
correctly in the
eardrum / forehead
Please check the
operation method
Hardware problem
Environment
temperature
beyond the scope
of 10°C~40°C
(50.0°F~104.0°F)
for Ear 15°C~40°C
(59.0°F~104.0°F) for
Forehead
Operation error, More
than 15 seconds
without test result in
heart rate mode
26
SOLUTIONS
Insert a new battery
Take out battery,
reinsert battery
correctly
Insert a new battery
Follow user's manual
to point at the eardrum
/ forehead.
Follow user's manual
for correct reading.
Consult with your
dealer
Condition the
thermometer until it
reaches environment
temperature 10°C~40°C
(50.0°F~104.0°F) for
Ear 15°C~40~ (59.0°F~
104.0°F) for Forehead
Please repeat the
operation, following
the instructions in user
manual