Safety precautions

EMS accepts no liability for direct or consequential injury or damage resulting from improper use, arising in
particular through the non-observance of the operating instructions, or improper preparation and
maintenance. The provision of technical documents does not imply any authorization for any adjustment,
repair or modification of this EMS product.
Danger - New or repaired handpieces or instruments must be sterilized before first use
Before first use, clean, disinfect and sterilize new or repaired handpieces and instruments which you wish to
use with the Piezon Master 400. Please refer to the information provided in the operating instructions.
Danger - Use for the intended purpose only
Failure to observe the operating instructions may result in the patient or user suffering serious injury or the
product being damaged, possibly beyond repair. Before using the product, make sure that you have studied
and understood the operating instructions. This also applies to any product used with the Piezon Master 400.
Danger - For use by qualified and trained personnel only
The Piezon Master 400 must not be used by other than properly qualified and trained personnel, in medical
facilities.
Caution - Always examine the unit and accessories for damage before commencing treatment
Always examine the unit and accessories for damage before commencing treatment A damaged Piezon
Master 400 or damaged accessories must not be used and must be replaced.
Danger - Do not install where there is a risk of explosion
The Piezon Master 400 is not intended for operation in presence of inflammable anaesthetics or gases.
Danger - Contraindications
Ultrasonic oscillations may prevent cardiac pacemakers functioning properly. Therefore, EMS recommend
that patients with a cardiac pacemaker should not be treated with the Piezon Master 400.
Caution - Check the safety status after repair
The safety status of the Piezon Master 400 and accessories must comply with the applicable standard for
this type of product after it has undergone any type of repair or modification. A report must be drawn up in
respect of every inspection, repair or modification.
Caution - This product needs special precautions regarding electromagnetic compatibility and needs to be
installed and put into service according to the section "Electromagnetic compatibility" provided in this
document.
Caution - Certain types of mobile telecommunication equipment could potentially interfere with this product. The separation
distances recommended in the section "Electromagnetic compatibility" of this document must be taken into account
Danger - This product should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary,
this product should be observed to verify normal operation in the configuration in which it will be used.
Danger - The use of accessories and cables other than those specified or sold by EMS as replacement parts, may result in
increased emissions or decreased immunity of this product.