3 Function
Brio comprises a hydraulic body assembly containing a manually operated release valve.
Operation of the release valve removes resistance to movement and allows adjustment of plantar
flexion and dorsiflexion. The hydraulic body assembly is connected to a female pyramid lower
housing via two pivot pins.
4 Maintenance
Maintenance must be carried out by competent personnel.
It is recommended that the following maintenance is carried out annually:
•
Check that all attachment screws are secure, clean and reassemble as necessary
The user should be advised:
Any changes in performance of this device must be reported to the practitioner.
Changes in performance may include:
•
Any unusual noise
•
Reduced ankle support (free movement with release button not depressed)
•
M-L movement or 'play' in pivots
The practitioner must also be informed of any changes in body weight and/or activity level.
Cleaning
Use a damp cloth and mild soap to clean the outside surfaces.
DO NOT use aggressive cleansers. Thoroughly dry before use.
5 Limitations on Use:
Intended Life:
A local risk assessment should be carried out based upon activity and usage.
Lifting Loads:
User weight and activity is governed by the stated limits.
Load carrying by the user should be based on a local risk assessment.
Environment:
Avoid exposing the product to corrosive elements such as water, acids and other
liquids. Also avoid abrasive environments such as those containing sand for example as
these may promote premature wear.
Exclusively for use between -15˚C to 50˚C.
It is recommended that only Blatchford products be used in conjunction with the product.