Autoclaving; Uv Sterilization - VITLAB micropipette-8 Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 63

11. Autoclaving

The pipette is completely autoclavable at 121 °C (250 °F), 2 bar
absolute (30 psi) with a holding time of at least 15 minutes, ac-
cording to DIN EN 285.
1. Eject the pipette tips.
2. Autoclave the complete pipette without any further disassem-
bling.
3. Allow the pipette to completely cool and dry.
Note:
The effectiveness of the autoclaving must be
verified by the user. Maximum reliability is obtained with vacu-
um sterilization. We recommend the use of sterilization bags.
Attention!
Prior to autoclaving, the volume adjustment
must be set on an available numbered volume (e.g., 11.25 or
11.26 but not between).
If the pipette is autoclaved frequently, the pistons should be oiled
with silicone oil in order to preserve smooth movement. Please
use only the recommended silicone oil, see accessories page 39. If
necessary after sterilization, tighten the connection between the
hand grip and the manifold.

12. UV sterilization

The unit can withstand the usual output of a UV sterilization
lamp. The effects of the UV may cause some color change.
32

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Micropipette-12

Tabla de contenido