POWER ON functions

Changing settings

This function enables you to permanently program limits and settings so they are
maintained even after switching the instrument ON/OFF. This results in a
customised instrument.
The setting can only be effected in the respective functions (central switch position).
STORE user defined Parameters:
a) Continuity Level
Switch to Ohm position.
Press button.
for continuity and afterwards
Big display shows
.
The small display shows the actual continuity trigger level in a range from 0.1
to 50.0Ω. Incrementing steps are 0.1Ω within the range of 0.2Ω up to 5Ω -
from 5Ω to 50Ω stepping is 1Ω. Change the continuity level with the
keyand select the digit with
.
Confirming this display with the
without saving.
b) Limit value programming
Switch to desired measuring function.
Press the
key- large display shows
The small display shows the LIMIT MIN value for the selected function (rotary switch
position). Use
skey to switch display to
function. Press
again to the show LIMIT
Change the LIMIT MIN with the buttons
The entry is completed by pressing the
store LIMIT MIN and exit.
Press the
key, the large display shows
between HI and LO limit).The small display shows the LIMIT MAX value or OFF
(same as above).
Change the LIMIT MIN with the buttons
The entry is completed by pressing the
If the instrument is in the setup mode it does not switch to power down (power
saving) mode.
.
key.
exits and saves.Turn rotary switch
.
. In this display
has no
(left most digit begins to flash) and.
button or pressing
button to
the (
key wraps arround
and.
button. Values are stored permanently.