Technical Data
Continuity check:
Buzzer activated for external resistance within 100ms
Buzzer level adjustable with LIMIT function from 0,2...50Ω.
Test lead compensation (ZERO) from 0.01...5Ω.
Voltage U measurement
DC: Input resistance:1MΩ / 100pF
Measuring
Resolution
range
2V
1mV
20V
10mV
200V
100mV
600V
1V
SMR
Serial mode rejection approx. 60dB at 50
and 60Hz for DC (influence of AC signals
to DCV-display)
CMR
Common mode rejection approx. 80dB at
50 and 60Hz
Auto range setting time:
1.5s
Max. voltage frequency product:
4
5x10
VHz
Over-range up to 1000V with flashing
AC: Specifications valid for AC RMS >5% of range
Measuring range Resolution
2V
1mV
20V
10mV
200V
100mV
600V
1V
Displayed values for voltage measurements: AC, DC and
AC+DC+F. Crest factor 3 for values at the end of the ranges
V
: 1200V
.
peak
P
Over-range up to 800V with flashing
Continuous logging and output through RS232 interface
possible.
Measuring procedure: The AC RMS and DC part are measured sequentially and
then the TRMS value calculated according to following formula:
Intrinsic error VDC
±(3% of rdg+10D)
±(2% of rdg+2D)
symbol.
Intrinsic error VAC
15Hz – 1kHz
±(3% of rdg+20D)
±(2% of rdg+5D)
±(2% of rdg+3D)
±(2% of rdg+2D)
symbol.
2
+
U
U
AC
DC
- 37 -
2