General specification
Storage temperature:
Storage relative humidity:
Index of protection:
Power supply: Qty.
Fuse (user replaceable):
Weight (minus batteries):
Possible number of operations during battery life:
When the following resistance is connected to a measuring
terminal alternating between 5 seconds loading and intervals
of 25 seconds, the number of measurements that it is
possible to make, until the battery voltage falls 6.6V will be:
250V Range
0.25MΩ approx.2400 times minimum
500V Range
0.5MΩ approx.1200 times minimum
1000V Range 1MΩ approx. 600 times minimum
Ωx 1 Range
1Ω approx. 800 times minimum
Applied Standards:
Operation: EN 61557-1/2/4
Safety:
EN61010, Cat.III 300V, Pollution degree 2
Protection: IEC60529
EMC:
IEC 801 & EN 55022
Accessories supplied:
Test-leads, Crocodile-clips, Test-lead pouch, Carrying strap
and Instructions.
-20℃ to 60℃
85% maximum
IP40
6 X 1.5V AA Alkaline batteries,
type R6 or equivalent.
500mA/600V, 1
/
x
/
1
1
"
4
4
(32 x 6.35mm) fast acting,
h i g h - b r e a k i n g c a p a c i t y ,
ceramic.
540g

4 Principles of Measurement

The ISO-TECH IIT-2301 Insulation and Continuity Tester
performs two basic electrical functions. As a continuity
tester the instrument can be used to measure low values of
resistance between two points in an electrical circuit. In this
mode the instrument acts as a low voltage current source. The
resistance is calculated from the measurement of the voltage
across and the current through the conductor and displayed
directly on the meter. To avoid measurement errors, ensure the
instrument has a good low impedance connection to the circuit
under test. Circuits connected in parallel with the circuit under
test may affect the accuracy of measurements.
Temporary
test at socket
link
between L and E
Typical arrangement for measuring the continuity of
protective conductors.
 Warning! Verify the circuit is de-energised before
testing.