Appendix 1: TECHNICAL SPECIFICATIONS
-Supply: 230 V/12v 50 Hz
-Power 25 W
- RO membrane capacity: 75 GPD*
-Tank capacity: Total 11L / available 8 L / min 1 L
-Recharge time:1 h depending on inlet water
-Size (width/length/height): 345mmx350mmx360mm
-Safety: Isolation class II
Working conditions:
-Inlet pressure 3-6 bar
-Inlet water temperature 4-45 °C
-Máx inlet TDS*: More than 1500 ppm contact technical service
-TDS* removal 95+/-1,5 %
-Maximum hardness 25 french degrees
-pH range 4-11
-Maximum production 280 l/d
-Estimated total production: 200 l/d, depending on inlet water
Sediment prefilter:
- Composition: Poliester
- Function: Retention of particles in suspension > 5µ
- Aprox. duration: 12 months depending on inlet water
- Maximum flow: 4l/m at 4 bar
Activated carbon prefilter:
- Composition: Activated charcoal
- Function: chlorine elimination
- Aprox. duration: 12 months. Depending on inlet water
- Chlorine retention: 99,9 %
Reverse osmosis membrane:
- Composition: Poliamide TFC
- Maximum capacity: 1x75 GPD* = 12 l./h depending on inlet water
- Aprox. duration: 3-5 years. Depending on inlet water. Depending on inlet water
Remineralizing postfilter
-Composition: carbon and algas fosilizadas
-Function: pH stabilization and water remineralization
- Aprox. duration: 3-5 years
UV-C disinfection unit
-Composition: UV-C lamp 15 W
-Function: Disinfection of water stored in the tank with germicide action
-Duration: 8.000 hours
*GPD – Galons per day. 1 galon = 3.78 litres
*TDS – Total disolved solids in water