St. Jude Medical MHV-HHR Instrucciones De Uso página 15

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 63
Step
Cycle
1
Pre-Wash
2
Alkaline
Detergent
Wash
3
Neutralizing
Rinse
4
Intermediate
Rinse
5
Disinfection
Rinse
6
Dry
After each cleaning cycle, inspect instruments for signs of cracking,
crazing, or degradation that may affect
Sterilization
Sterilize instruments before each use.
2. pH10.4 – 10.8 (2 – 10 ml/l, determined in deionized water, 20°C). In tap water or
softened water as well as with carried-over pre-rinsing water, the pH value may vary -
Neodisher MediClean forte Technical Data Sheet, 05/2015.
Duration
Requirements
2 minutes
Cold tap water
5 minutes
55-60°C / 131-
140°F Neodisher™
MediClean forte
diluted to 0.5% (v/v)
5 minutes
~25°C tap water
4 minutes
Cold tap water
5 minutes
93°C tap water
--
--
9
Cycle
Do not use
detergents or
additives. Drain
when complete.
Drain when
complete.
,
2
Drain when
complete.
Drain when
complete.
Drain when
complete.
Allow to air dry
function.2