Note: The following tests must be performed after a load cell has been replaced:
Check side load and adjust as needed
Calibrate with maximum load
9.4
Replacing Levelling Feet
1)
Stand scale up
2)
Unhook retaining springs from the defective foot and remove it
3)
Lubricate new foot with O-ring and insert into the load cells
4)
Hook retaining springs into the new foot
EN-9