Proceed as follows to set the sensor type:
You will see the start screen after the programs has loaded.
Select „Service Programs" with the cursor up key (o) and confirm
this with the Enter key (E).
Fig. 26: Start screen
Then select „After-Sale Service" with the cursor up key (o) and
confirm this likewise with the Enter key (E).
Fig. 27: „After-Sale Service" selection
Now, you have the option to make the required settings through
„Setup". Confirm this with the Enter key (E).
Fig. 28: „Service" selection
1 689 978 435N
08.12.08
Make the following selection with the cursor keys:
WA 900:
„Old Sensor"
WA 920:
"Radio Device"
Confirm your selection with the Enter key (E).
Fig. 29: Sensor selection
After pressing the <Esc> key twice, you are asked to confirm your
selection.
Fig. 30: Confirm the selection
The settings take effect after this and the wheel aligner is ready
for use.
In operating instructions 1 689 979 873N you can find a detail
functional and operational description of the wheel aligner.
3.3
Maintenance notes
- Do not use abrasive powders for cleaning, nor any cleaners that
can dissolve plastics. Ensure that no liquids get inside the unit.
- The unit must be switched off completely before making any
changes to the settings or carrying out maintenance work.
- Do not remove any parts of the unit except for servicing work.
- Ensure that the working area is kept clean. Under no circum-
stances should you use compressed air or a water jet to remove
dirt or foreign bodies. Ensure that dust is neither created nor
blown about during the cleaning work.
- Handle the sensors with care and keep them in a dry place. This
avoids compensation errors and false measurements.
- Calibrate the sensors every 6 months.
- Keep the glass covers of the sensors in a clean condition. Use
a soft cloth for cleaning. Never remove or scratch the glass
covers.
- Keep the guides of the clamps clean.
- The turn and slip plates must always be kept clean and may not
be oiled or greased.
29