is intended for use in the electromagnetic environment specified below. The costumer or the user of
Below mentioned tests are applicable to enclosure. Additionally, test EN-IEC 61000-4-2 is applicable to input/output signal and patient connections.
Immunity test
Electrostatic discharge (ESD)
EN-IEC 61000-4-2
Radiated RF
EN-IEC 61000-4-3
Proximity fields from RF
communications
EN-IEC 61000-4-3
Power frequency (50 / 60 Hz)
magnetic field
EN-IEC 61000-4-8
EN 23
Guidance and manufacturer's declaration – electromagnetic immunity
EN-IEC 60601 test level
Compliance level
±8 kV contact
±8 kV contact
±2, ±4, ±8, ±15 kV,
air
±2, ±4, ±8, ±15 kV, air
10 V/m
10 V/m
80 MHz to 2.7 GHz
80% AM at 1kHz
(View following table).
30 A/m
30 A/m
somnibel
Somnibel
should assure that it is used in such an environment.
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should
be used no closer than 30cm to any part of
cables. The distance is calculated according to:
where P is the output power rating of the transmitter in watts
(W), d is the recommended separation in meters (m) and E is the
immunity level in V/m.
Only applicable to devices sensitive to magnetic fields.
User's Manual
Somnibel
including
534-800-MUM_Rev.1.06