CLEANING & DISINFECTING
CAUTION
or disinfecting the monitor. The screen can be scratched, permanently damaging the device.
CAUTION
using the approved low‑temperature processes recommended by Verathon
WARNING
or body fluids capable of transmitting pathogens, all cleaning facilities must be in compliance
with (U.S.) OSHA Standard 29  CFR  1910.1030 "Bloodborne Pathogens" or an equivalent
standard. For more information, visit www.osha.gov.
WARNING
handling and disposing of cleaning, disinfection, or sterilization solutions.
WARNING
Reuse, reprocessing, or resterilization may create a risk of contamination of the device.
WARNING
Single‑use blades arrive sterilized by ethylene oxide, and they do not require cleaning,
disinfection, or sterilization prior to use. Dispose of single‑use blades once they have been
used. Do not attempt to disinfect and reuse single‑use video laryngoscopes.
Table 3
describes the risk assessment for each system component, including the Spaulding
classification for the minimum required disinfection level. Prior to each use, ensure that
each system component has been cleaned, disinfected, or sterilized according to the
guidance provided in this table.
Table 3.
System Risk Classification
Device
Monitor
Charging cradle
Single‑use blades
Video Baton 2.0
GVL
Stats
®
GlideRite
Rigid Stylet
®
GlideRite Single‑Use Stylet
*
Clean the charging cradle when visibly soiled and on a regular basis.
For instructions, see the GlideRite Rigid Stylet Operations and Maintenance Manual.
Single‑use blades and stylets may not be reused. Dispose of single‑use items after use.
§
||
Video batons require low‑level disinfection between uses, but may be high‑level disinfected if desired.
X
Checked boxes show minimum disinfection level requirement.
Shaded areas indicate that the disinfection or sterilization level is not required or not compatible with the device.
Unshaded areas show permissible levels of disinfection or sterilization based on compatibility with the device materials.
Ensure that you do not use any abrasive tools when cleaning
This product may only be cleaned, disinfected, or sterilized by
Because the product may be contaminated with human blood
Ensure that you follow the manufacturer's instructions for
Do not reuse, reprocess, or resterilize single‑use components.
The monitor must be cleaned before initial use.
Use
Reusable
Reusable
*
Single‑use
§
Reusable
||
Single‑use
§
Reusable
Single‑use
§
Spaulding
Clean
Classification
Non‑critical
X
Non‑critical
Non‑critical
Semi‑critical
.
®
Disinfect
Sterile
Low
High
X
X
13