ELECTROMAGNETIC EMISSIONS
Table 11. Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The system is intended for use in the electromagnetic environment specified below. The customer or
the user of the system should ensure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000‑3‑2
Voltage fluctuations/flicker emissions
IEC 61000‑3‑3
RECOMMENDED SEPARATION DISTANCES
Table 12. Recommended Separation Distances between Portable and Mobile RF Communications
Equipment and the System
The system is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the system can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the system as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
ACCESSORY CONFORMANCE TO STANDARDS
To maintain electromagnetic interference (EMI) within certified limits, the system must be
used with the cables, components, and accessories specified or supplied by Verathon
additional information, see
other than those specified or supplied may result in increased emissions or decreased
immunity of the system.
Table 13.
EMC Standards for Accessories
Accessory
Monitor power adapter
Charging cradle power adapter
24
Compliance
Group 1
Class A
Class A
Complies
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d=1.2 √P
0.12
0.38
1.2
3.8
12
System Parts &
Accessories. The use of accessories or cables
Electromagnetic Environment – Guidance
The system uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The system is suitable for use in all
establishments other than domestic and those
directly connected to the public low‑voltage
power supply network that supplies buildings
used for domestic purposes.
80 MHz to 800 MHz
d=1.2 √P
0.12
0.38
1.2
3.8
12
Max Cable Length
1.5 m (4.9 ft)
1.5 m (4.9 ft)
800 MHz to 2.5 GHz
d=2.3 √P
0.23
0.73
2.3
7.3
23
. For
®