DeVilbiss Healthcare intellipap CPAP Serie Manual De Instrucciones página 19

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Display Example
P : 1 0 . 0 S
T R R : 1 5
I / E : 1 : 1 . 0 ( 5 0 % )
L K : 1 2 3 S
S E V : A S
M V : 1 2 3 . 4
V T : 1 2 3 4
IPAP
10.0 cmH2O r 5.0
EPAP
5.0 cmH2O r 5.0
Set Pressure
10.0 cmH2O
Delay Time
45 minutes
Delay Pressure
5.0 cmH2O
IPAP Rounding
0 (1234 ms)
EPAP Rounding
0 (1234 ms)
Insp. Trigger
5
Exp. Trigger
5
Backup Rate
15 BPM
I/E Ratio
1 : 1.0 (50%)
Tubing Length
1.8 m (6 ft)
Enable Menu
Press 
A-DV56
Description
Real time info screen, top row shows measured pressure (P), Trigger
source (Spontaneous or Timed), breath state (dash in center of top
row goes high during inhale and low during exhale). Bottom row
shows I:E ratio and % inhale.
Real time info screen, top row shows Leak rate (LK), Trigger source
(Spontaneous or Timed), breath state (dash in center of top row goes
high during inhale and low during exhale), Event Markers (EV),
Bottom row shows Minute Volume (MV) and Tidal Volume (VT)
This is a prescription setting which controls the pressure during inhale,
this also shows the pressure delta (5.0 in example) between inhale
and exhale pressure settings
This is a prescription setting which controls the pressure during
exhale, this also shows the pressure delta (5.0 in example) between
inhale and exhale pressure settings
This is a prescription setting which controls the operating pressure.
Selectable delay time setting. Your provider has set the delay time.
This is a prescription setting which controls the operating pressure
during the delay time.
This is a setting which controls the comfort rounding when switching to
IPAP pressure, settings are 0 thru 5 also shows measured rise time of
the last breath in milliseconds.
This is a setting which controls the comfort rounding when switching to
EPAP pressure, settings are 0 thru 5 also shows measured rise time
of the last breath in milliseconds.
This is a setting which controls the sensitivity of the inspiratory trigger,
can be set from 1 to 10, 1 is most sensitive, 10 is least sensitive.
This is a setting which controls the sensitivity of the expiratory trigger,
can be set from 1 to 10, 1 is most sensitive, 10 is least sensitive.
This is a setting which affects the backup pressure cycle rate. A new
breath will be triggered if the patient's breathing is slower than this
setting.
This is a setting which affects the back up pressure cycle rate. This is
the ratio of the inhale to exhale time and is expressed as 1:X. A new
exhale will be triggered by the Bilevel if the patient's breathing is
slower than this setting, also linked to Backup Rate. % inhale is shown
in parenthesis.
The standard tubing length supplied with the device is 6 feet (1.8
meters). However, you can adjust your device to accommodate an air
supply tubing length of 10 feet (3.0 meters). Replacement tubing can
be obtained through your home care provider (part numbers listed at
the end of this manual).
Pressing the down key enters the Enable menu. Refer to next section
for Enable Menu details.
Mode
All
Bilevel
Modes
All
Bilevel
Modes
All
Bilevel
Modes
All
Bilevel
Modes
CPAP Mode
All
CPAP Mode
All
Bilevel
Modes
All
Bilevel
Modes
Bilevel
S & ST
Modes
Bilevel
S & ST
Modes
Bilevel
ST & T
Modes
Bilevel
ST & T
Modes
All
All
eN - 19

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido