Descargar Imprimir esta página

wtw pH 330i Instrucciones De Operación página 10

Ph/ mv medidor
Ocultar thumbs Ver también para pH 330i:

Publicidad

Idiomas disponibles

Idiomas disponibles

Calibration
pH 330i/340i
pH
flf" I I
1
|""j ~| ~j
II
°C
.•H
-in
MJ
C
pH
i l l
IJQ
I.OX
n
J L U
°C
ConCal procedure:
• Connect the pH electrode to the measuring instrument.
• Press the <CAL> key repeatedly until the display ASYand the function
display ConCal appears. The sensor symbol displays the evaluation of
the last calibration (or no sensor symbol in the delivery state or after the
measurement parameter has been reset).
• Immerse the pH electrode in the first buffer solution
(pH 7.0 ± 0.5 in two-point calibration).
• When measuring without a temperature sensor:
Enter the temperature of the first buffer solution
with <A> <T> while pressing the <RUN/ENTER> key.
• Press the <RUN/ENTER> key.
The measured pH value appears on the display.
• Set the measured value to the nominal pH value of the buffer solution (at
the current temperature) with <A> <•>.
• If the measured value is stable, press the <RUN/ENTER> key. The value
of the asymmetry appears.
• Press the <RUN/ENTER> key. SLO appears.
Note
At this point, the ConCal calibration can be terminated with
<M>. This corresponds to a single-point calibration. To do
this, the instrument uses the Nernst slope (-59.16 mV/pH at
25 °C) and determines the asymmetry of the electrode.
pH
_ p _
_
re
• Immerse the two-point calibration pH electrode in the second buffer so-
lution to continue.
• When measuring without a temperature sensor:
Enter the temperature of the second buffer solution with <A> <T> while
pressing the <RUN/ENTER> key.
• Press <RUN/ENTER>.
• Set the measured value to the nominal pH value of the buffer solution (at
the current temperature) with •
pH
pH
mV/
When the measured value is stable, press the <RUN/ENTER> key. The
instrument displays the value of the slope and the calibration evaluation.
Press <RUN/ENTER>. The instrument displays the value of the asym-
m e t f
y-
Switch to the measuring mode with <M>.
10

Publicidad

loading

Este manual también es adecuado para:

Ph 340i