TCM Endo III
8)
5.1 The TCM Endo III unit
Starting the device:
Press green "ON/OFF" switch (rear of unit).
Display:
The digital display indicates the current drill speed or the selected torque level.
Keypad/Control functions:
„Motor":
Switch On/Off for micromotor (green light illuminates when the motor is in operation)
„▲":
Speed up or increase torque
„▼":
Speed down or reduce torque
To change motor direction of rotation → a warning tone will sound when the motor
„Rev/Forw":
operates in reverse mode
„1:1":
Press this key if using handpiece 1:1
Speed : from 1,200 to 16,000 rpm
Torque: 1, 2, 3, 4, 5, 7, 10, 13, 16, 19 Nmm
„8:1":
Press this key if using contra angle 8:1
Speed: from 150 to 2,000 rpm
Torque: 2, 4, 6, 10, 15, 20, 25, 30, 40, 50 Nmm
„16:1":
Press this key if using contra angle 16:1
Speed: from 75 to 1'000 Upm
Torque: 5, 10, 15, 20, 25, 30, 40, 50, 75, 100 Nmm
„ATC":
AL- and AP- mode selection for desired torque function
The „Motor" key functions only when the foot switch is not connected
4
5 Operation
NOUVAG AG
No. 31817 29/05